Poor reproducibility of compression elastography in the Achilles tendon: same day and consecutive day measurements.
OBJECTIVE
To determine the reproducibility of compression elastography (CE) when measuring strain data, a measure of stiffness of the human Achilles tendon in vivo, over consecutive measures, consecutive days and when using different foot positions.
MATERIALS AND METHODS
Eight participants (4 males, 4 females; mean age 25.5 ± 2.51 years, range 21-30 years; height 173.6 ± 11.7 cm, range 156-189 cm) had five consecutive CE measurements taken on one day and a further five CE measures taken, one per day, at the same time of day, every day for a consecutive 5-day period. These 80 measurements were used to assess both the repeatability and reproducibility of the technique. Means, standard deviations, coefficient of variation (CV), Pearson correlation analysis (R) and intra-class correlation coefficients (ICC) were calculated.
RESULTS
For CE data, all CVs were above 53%, R values indicated no-to-weak correlations between measures at best (range 0.01-0.25), and ICC values were all classified in the poor category (range 0.00-0.11). CVs for length and diameter measures were acceptably low indicating a high level of reliability.
CONCLUSIONS
Given the wide variation obtained in the CE results, it was concluded that CE using this specific system has a low level of reproducibility for measuring the stiffness of the human Achilles tendon in vivo over consecutive days, consecutive measures and in different foot positions
Do quantitative and qualitative shear wave elastography have a role in evaluating musculoskeletal soft tissue masses?
Objectives: To determine if quantitative and qualitative shear wave elastography have roles in evaluating musculoskeletal masses. Methods: 105 consecutive patients, prospectively referred for biopsy within a specialist sarcoma centre, underwent B-mode, quantitative (m/s) and qualitative (colour map) shear wave elastography. Reference was histology from subsequent biopsy or excision where possible. Statistical modelling was performed to test elastography data and/or B-mode imaging in predicting malignancy. Results: Of 105 masses, 39 were malignant and 6 had no histology but benign characteristics at 12 months. Radiologist agreement for B-mode and elastography was moderate to excellent Kw 0.52-0.64; PABAKw 0.85-0.90). B-Mode imaging had 78.8% specificity, 76.9% sensitivity for malignancy. Quantitatively, adjusting for age, B-mode and lesion volume there was no statistically significant association between longitudinal velocity and malignancy (OR [95% CI] 0.40[0.10, 1.60], p=0.193), but some evidence that higher transverse velocity was associated with decreased odds of malignancy (0.28[0.06, 1.28], p=0.101). Qualitatively malignant masses tended to be towards the blue spectrum (lower velocities); 39.5% (17/43) of predominantly blue masses were malignant, compared to 14.3% (1/7) of red lesions. Conclusions: Quantitatively and qualitatively there is no statistically significant association between shear wave velocity and malignancy. There is no clear additional role to B-mode imaging currently. Key Points: • Correlation between shear wave velocity and soft tissue malignancy was statistically insignificant• B-mode ultrasound is 76.9 % sensitive and 78.8 % specific• Statistical models show elastography does not significantly add to lesion assessmen
New sonographic measures of peripheral nerves: a tool for the diagnosis of peripheral nerve involvement in leprosy
The EFSUMB Guidelines and Recommendations for the Clinical Practice of Elastography in Non-Hepatic Applications: Update 2018
This manuscript describes the use of ultrasound elastography, with the exception of liver applications, and represents an update of the 2013 EFSUMB (European Federation of Societies for Ultrasound in Medicine and Biology) Guidelines and Recommendations on the clinical use of elastography
Surgery for greater trochanteric pain syndrome after total hip replacement confers a poor outcome
Old and new oral anticoagulants : food, herbal medicines and drug interactions
The most commonly prescribed oral anticoagulants worldwide are the vitamin K antagonists (VKAs) such as warfarin. Factors affecting the pharmacokinetics of VKAs are important because deviations from their narrow therapeutic window can result in bleedings due to over-anticoagulation or thrombosis because of under-anticoagulation. In addition to pharmacodynamic interactions (e.g., augmented bleeding risk for concomitant use of NSAIDs), interactions with drugs, foods, herbs, and over-the-counter medications may affect the risk/benefit ratio of VKAs. Direct oral anticoagulants (DOACs) including Factor Xa inhibitors (rivaroxaban, apixaban and edoxaban) and thrombin inhibitor (dabigatran) are poised to replace warfarin. Phase-3 studies and real-world evaluations have established that the safety profile of DOACs is superior to those of VKAs. However, some pharmacokinetic and pharmacodynamic interactions are expected. Herein we present a critical review of VKAs and DOACs with focus on their potential for interactions with drugs, foods, herbs and over-the-counter medications
The time-averaged inflammatory disease activity estimates the progression of erosions in MRI of the sacroiliac joints in ankylosing spondylitis
Enthesiopathy of the flexor carpi ulnaris at the pisiform: Findings of high-frequency sonography
- …