561 research outputs found
Ion source for tests of ion behavior in the KATRIN beam line
An electron-impact ion source based on photoelectron emission was developed
for ionization of gases at pressures below 1e-4 mbar in an axial magnetic field
in the order of 5 T. The ion source applies only DC fields, which makes it
suitable for use in the presence of equipment sensitive to radio-frequency (RF)
fields. The ion source was succesfully tested under varying conditions
regarding pressure, magnetic field and magnetic-field gradient, and the results
were studied with the help of simulations. The processes in the ion source are
well understood and possibilities for further optimization of generated ion
currents are clarified.Comment: 10 pages, 13 figure
Monitoring of tritium purity during long-term circulation in the KATRIN test experiment LOOPINO using laser Raman spectroscopy
The gas circulation loop LOOPINO has been set up and commissioned at Tritium
Laboratory Karlsruhe (TLK) to perform Raman measurements of circulating tritium
mixtures under conditions similar to the inner loop system of the neutrino-mass
experiment KATRIN, which is currently under construction. A custom-made
interface is used to connect the tritium containing measurement cell, located
inside a glove box, with the Raman setup standing on the outside. A tritium
sample (purity > 95%, 20 kPa total pressure) was circulated in LOOPINO for more
than three weeks with a total throughput of 770 g of tritium. Compositional
changes in the sample and the formation of tritiated and deuterated methanes
CT_(4-n)X_n (X=H,D; n=0,1) were observed. Both effects are caused by hydrogen
isotope exchange reactions and gas-wall interactions, due to tritium {\beta}
decay. A precision of 0.1% was achieved for the monitoring of the T_2
Q_1-branch, which fulfills the requirements for the KATRIN experiment and
demonstrates the feasibility of high-precision Raman measurements with tritium
inside a glove box
Recommended from our members
Improved Upper Limit on the Neutrino Mass from a Direct Kinematic Method by KATRIN.
We report on the neutrino mass measurement result from the first four-week science run of the Karlsruhe Tritium Neutrino experiment KATRIN in spring 2019. Beta-decay electrons from a high-purity gaseous molecular tritium source are energy analyzed by a high-resolution MAC-E filter. A fit of the integrated electron spectrum over a narrow interval around the kinematic end point at 18.57 keV gives an effective neutrino mass square value of (-1.0_{-1.1}^{+0.9}) eV^{2}. From this, we derive an upper limit of 1.1 eV (90% confidence level) on the absolute mass scale of neutrinos. This value coincides with the KATRIN sensitivity. It improves upon previous mass limits from kinematic measurements by almost a factor of 2 and provides model-independent input to cosmological studies of structure formation
Commissioning of the vacuum system of the KATRIN Main Spectrometer
The KATRIN experiment will probe the neutrino mass by measuring the
beta-electron energy spectrum near the endpoint of tritium beta-decay. An
integral energy analysis will be performed by an electro-static spectrometer
(Main Spectrometer), an ultra-high vacuum vessel with a length of 23.2 m, a
volume of 1240 m^3, and a complex inner electrode system with about 120000
individual parts. The strong magnetic field that guides the beta-electrons is
provided by super-conducting solenoids at both ends of the spectrometer. Its
influence on turbo-molecular pumps and vacuum gauges had to be considered. A
system consisting of 6 turbo-molecular pumps and 3 km of non-evaporable getter
strips has been deployed and was tested during the commissioning of the
spectrometer. In this paper the configuration, the commissioning with bake-out
at 300{\deg}C, and the performance of this system are presented in detail. The
vacuum system has to maintain a pressure in the 10^{-11} mbar range. It is
demonstrated that the performance of the system is already close to these
stringent functional requirements for the KATRIN experiment, which will start
at the end of 2016.Comment: submitted for publication in JINST, 39 pages, 15 figure
The KATRIN Pre-Spectrometer at reduced Filter Energy
The KArlsruhe TRItium Neutrino experiment, KATRIN, will determine the mass of
the electron neutrino with a sensitivity of 0.2 eV (90% C.L.) via a measurement
of the beta-spectrum of gaseous tritium near its endpoint of E_0 =18.57 keV. An
ultra-low background of about b = 10 mHz is among the requirements to reach
this sensitivity. In the KATRIN main beam-line two spectrometers of MAC-E
filter type are used in a tandem configuration. This setup, however, produces a
Penning trap which could lead to increased background. We have performed test
measurements showing that the filter energy of the pre-spectrometer can be
reduced by several keV in order to diminish this trap. These measurements were
analyzed with the help of a complex computer simulation, modeling multiple
electron reflections both from the detector and the photoelectric electron
source used in our test setup.Comment: 22 pages, 12 figure
Final Results from phase II of the Mainz Neutrino Mass Search in Tritium Decay
The paper reports on the improved Mainz experiment on tritum
spectroscopy which yields a 10 times' higher signal to background ratio than
before. The main experimental effects and systematic uncertainties have been
investigated in side experiments and possible error sources have been
eliminated. Extensive data taking took place in the years 1997 to 2001. A
residual analysis of the data sets yields for the square of the electron
antineutrino mass the final result of eV/c. We derive an upper limit of
eV/c at 95% confidence level for the mass itself.Comment: 22 pages, 22 figures submitted to EPJ
- …