Determination of the map of efficiency of the J-PET detector with the GATE package
A novel PET detector consisting of strips of polymer scintillators is being
developed by the J-PET Collaboration. The map of efficiency and the map of
geometrical acceptance of the 2-strip J-PET scanner are presented. Map of
efficiency was determined using the Monte Carlo simulation software GATE based
on GEANT4. Both maps were compared using method based on the chi2 test.Comment: 14 pages, 9 figures, proceeding from conference Symposium on Positron
Emission Tomography: http://koza.if.uj.edu.pl/pet-symposium-2013
Novel method for hit-position reconstruction using voltage signals in plastic scintillators and its application to Positron Emission Tomography
Currently inorganic scintillator detectors are used in all commercial Time of
Flight Positron Emission Tomograph (TOF-PET) devices. The J-PET collaboration
investigates a possibility of construction of a PET scanner from plastic
scintillators which would allow for single bed imaging of the whole human body.
This paper describes a novel method of hit-position reconstruction based on
sampled signals and an example of an application of the method for a single
module with a 30 cm long plastic strip, read out on both ends by Hamamatsu
R4998 photomultipliers. The sampling scheme to generate a vector with samples
of a PET event waveform with respect to four user-defined amplitudes is
introduced. The experimental setup provides irradiation of a chosen position in
the plastic scintillator strip with an annihilation gamma quanta of energy
511~keV. The statistical test for a multivariate normal (MVN) distribution of
measured vectors at a given position is developed, and it is shown that signals
sampled at four thresholds in a voltage domain are approximately normally
distributed variables. With the presented method of a vector analysis made out
of waveform samples acquired with four thresholds, we obtain a spatial
resolution of about 1 cm and a timing resolution of about 80 p
Analysis framework for the J-PET scanner
J-PET analysis framework is a flexible, lightweight, ROOT-based software
package which provides the tools to develop reconstruction and calibration
procedures for PET tomography. In this article we present the implementation of
the full data-processing chain in the J-PET framework which is used for the
data analysis of the J-PET tomography scanner. The Framework incorporates
automated handling of PET setup parameters' database as well as high level
tools for building data reconstruction procedures. Each of these components is
briefly discussed.Comment: 6 pages, 1 figur
A pilot study of the novel J-PET plastic scintillator with 2-(4-styrylphenyl)benzoxazole as a wavelength shifter
For the first time a molecule of 2-(4-styrylphenyl)benzoxazole containing
benzoxazole and stilbene groups is applied as a scintillator dopant acting as a
wavelength shifter. In this article a light yield of the plastic scintillator,
prepared from styrene doped with 2 wt% of 2,5-diphenylbenzoxazole and 0.03 wt%
of 2-(4-styrylphenyl)benzoxazole, is determined to be as large as 60% 2%
of the anthracene light output. There is a potential to improve this value in
the future by the optimization of the additives concentrations
PALS investigations of free volumes thermal expansion of J-PET plastic scintillator synthesized in polystyrene matrix
The polystyrene dopped with 2,5-diphenyloxazole as a primary fluor and
2-(4-styrylphenyl)benzoxazole as a wavelength shifter, prepared as a plastic
scintillator was investigated using positronium probe in wide range of
temperatures from 123 to 423 K. Three structural transitions at 260 K, 283 K
and 370 K were found in the material. In the o-Ps intensity dependence on
temperature, the significant hysteresis is observed. Heated to 370 K, the
material exhibits the o-Ps intensity variations in time.Comment: in Nukleonika 201
A novel method for calibration and monitoring of time synchronization of TOF-PET scanners by means of cosmic rays
All of the present methods for calibration and monitoring of TOF-PET scanner
detectors utilize radioactive isotopes such as e.g. Na or Ge,
which are placed or rotate inside the scanner. In this article we describe a
novel method based on the cosmic rays application to the PET calibration and
monitoring methods. The concept allows to overcome many of the drawbacks of the
present methods and it is well suited for newly developed TOF-PET scanners with
a large longitudinal field of view. The method enables also monitoring of the
quality of the scintillator materials and in general allows for the continuous
quality assurance of the PET detector performance.Comment: 10 pages, 7 figure
- …