251 research outputs found
Planck LFI flight model feed horns
this paper is part of the Prelaunch status LFI papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jinst The Low Frequency
Instrument is optically interfaced with the ESA Planck telescope through 11
corrugated feed horns each connected to the Radiometer Chain Assembly (RCA).
This paper describes the design, the manufacturing and the testing of the
flight model feed horns. They have been designed to optimize the LFI optical
interfaces taking into account the tight mechanical requirements imposed by the
Planck focal plane layout. All the eleven units have been successfully tested
and integrated with the Ortho Mode transducers.Comment: This is an author-created, un-copyedited version of an article
accepted for publication in JINST. IOP Publishing Ltd is not responsible for
any errors or omissions in this version of the manuscript or any version
derived from it. The definitive publisher authenticated version is available
online at 10.1088/1748-0221/4/12/T1200
The Planck-LFI flight model composite waveguides
The Low Frequency Instrument on board the PLANCK satellite is designed to
give the most accurate map ever of the CMB anisotropy of the whole sky over a
broad frequency band spanning 27 to 77 GHz. It is made of an array of 22
pseudo-correlation radiometers, composed of 11 actively cooled (20 K) Front End
Modules (FEMs), and 11 Back End Modules (BEMs) at 300K. The connection between
the two parts is made with rectangular Wave Guides. Considerations of different
nature (thermal, electromagnetic and mechanical), imposed stringent
requirements on the WGs characteristics and drove their design. From the
thermal point of view, the WG should guarantee good insulation between the FEM
and the BEM sections to avoid overloading the cryocooler. On the other hand it
is essential that the signals do not undergo excessive attenuation through the
WG. Finally, given the different positions of the FEM modules behind the focal
surface and the mechanical constraints given by the surrounding structures,
different mechanical designs were necessary. A composite configuration of
Stainless Steel and Copper was selected to satisfy all the requirements. Given
the complex shape and the considerable length (about 1.5-2 m), manufacturing
and testing the WGs was a challenge. This work deals with the development of
the LFI WGs, including the choice of the final configuration and of the
fabrication process. It also describes the testing procedure adopted to fully
characterize these components from the electromagnetic point of view and the
space qualification process they underwent. Results obtained during the test
campaign are reported and compared with the stringent requirements. The
performance of the LFI WGs is in line with requirements, and the WGs were
successfully space qualified.Comment: this paper is part of the Prelaunch status LFI papers published on
JINST: http://www.iop.org/EJ/journal/-page=extra.proc5/jins
Planck-LFI radiometers' spectral response
The Low Frequency Instrument (LFI) is an array of pseudo-correlation
radiometers on board the Planck satellite, the ESA mission dedicated to
precision measurements of the Cosmic Microwave Background. The LFI covers three
bands centred at 30, 44 and 70 GHz, with a goal bandwidth of 20% of the central
frequency.
The characterization of the broadband frequency response of each radiometer
is necessary to understand and correct for systematic effects, particularly
those related to foreground residuals and polarization measurements. In this
paper we present the measured band shape of all the LFI channels and discuss
the methods adopted for their estimation. The spectral characterization of each
radiometer was obtained by combining the measured spectral response of
individual units through a dedicated RF model of the LFI receiver scheme.
As a consistency check, we also attempted end-to-end spectral measurements of
the integrated radiometer chain in a cryogenic chamber. However, due to
systematic effects in the measurement setup, only qualitative results were
obtained from these tests. The measured LFI bandpasses exhibit a moderate level
of ripple, compatible with the instrument scientific requirements.Comment: 16 pages, 9 figures, this paper is part of the Prelaunch status LFI
papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jins
High Performances Corrugated Feed Horns for Space Applications at Millimetre Wavelengths
We report on the design, fabrication and testing of a set of high performance
corrugated feed horns at 30 GHz, 70 GHz and 100 GHz, built as advanced
prototypes for the Low Frequency Instrument (LFI) of the ESA Planck mission.
The electromagnetic designs include linear (100 GHz) and dual shaped (30 and 70
GHz) profiles. Fabrication has been achieved by direct machining at 30 GHz, and
by electro-formation at higher frequencies. The measured performances on side
lobes and return loss meet the stringent Planck requirements over the large
(20%) instrument bandwidth. Moreover, the advantage in terms of main lobe shape
and side lobes levels of the dual profiled designs has been demonstrated.Comment: 16 pages, 7 figures, accepted for publication in Experimental
Astronom
A multichannel reflectometer for edge density profile measurements at the ICRF antenna in ASDEX upgrade
A multichannel reflectometer will be built for the new three-straps ICRF antenna of ASDEX Upgrade (AUG), to study the density behavior in front of it. Ten different accesses to the plasma are available for the three reflectometer channels that can be interchanged without breaking the machine vacuum. Frequency is scanned from 40 GHz to 68 GHz, in 10 mu s, which corresponds to a cut-off density ranging from 10(18) divided by 10(19)m(-3) in the Right cut-off of the X-mode propagation, for standard toroidal magnetic field values of AUG
The linearity response of the Planck-LFI flight model receivers
In this paper we discuss the linearity response of the Planck-LFI receivers,
with particular reference to signal compression measured on the 30 and 44 GHz
channels. In the article we discuss the various sources of compression and
present a model that accurately describes data measured during tests performed
with individual radiomeric chains. After discussing test results we present the
best parameter set representing the receiver response and discuss the impact of
non linearity on in-flight calibration, which is shown to be negligible.Comment: this paper is part of the Prelaunch status LFI papers published on
JINST: http://www.iop.org/EJ/journal/-page=extra.proc5/jinst; This is an
author-created, un-copyedited version of an article accepted for publication
in JINST. IOP Publishing Ltd is not responsible for any errors or omissions
in this version of the manuscript or any version derived from it. The
definitive publisher authenticated version is available online at
10.1088/1748-0221/4/12/T12011
Thermal susceptibility of the Planck-LFI receivers
This paper is part of the Prelaunch status LFI papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jinst .
This paper describes the impact of the Planck Low Frequency Instrument front
end physical temperature fluctuations on the output signal. The origin of
thermal instabilities in the instrument are discussed, and an analytical model
of their propagation and impact on the receivers signal is described. The
experimental test setup dedicated to evaluate these effects during the
instrument ground calibration is reported together with data analysis methods.
Finally, main results obtained are discussed and compared to the requirements.Comment: This is an author-created, un-copyedited version of an article
accepted for publication in Journal of Instrumentation. IOP Publishing Ltd is
not responsible for any errors or omissions in this version of the manuscript
or any version derived from it. The definitive publisher authenticated
version is available online at 10.1088/1748-0221/4/12/T1201
Planck-LFI radiometers tuning
"This paper is part of the Prelaunch status LFI papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jinst"
This paper describes the Planck Low Frequency Instrument tuning activities
performed through the ground test campaigns, from Unit to Satellite Levels.
Tuning is key to achieve the best possible instrument performance and tuning
parameters strongly depend on thermal and electrical conditions. For this
reason tuning has been repeated several times during ground tests and it has
been repeated in flight before starting nominal operations. The paper discusses
the tuning philosophy, the activities and the obtained results, highlighting
developments and changes occurred during test campaigns. The paper concludes
with an overview of tuning performed during the satellite cryogenic test
campaign (Summer 2008) and of the plans for the just started in-flight
calibration.Comment: This is an author-created, un-copyedited version of an article
accepted for publication in JINST. IOP Publishing Ltd is not responsible for
any errors or omissions in this version of the manuscript or any version
derived from it. The definitive publisher authenticated version is available
online at http://dx.doi.org/10.1088/1748-0221/4/12/T12013
Planck-LFI: Design and Performance of the 4 Kelvin Reference Load Unit
The LFI radiometers use a pseudo-correlation design where the signal from the
sky is continuously compared with a stable reference signal, provided by a
cryogenic reference load system. The reference unit is composed by small
pyramidal horns, one for each radiometer, 22 in total, facing small absorbing
targets, made of a commercial resin ECCOSORB CR (TM), cooled to approximately
4.5 K. Horns and targets are separated by a small gap to allow thermal
decoupling. Target and horn design is optimized for each of the LFI bands,
centered at 70, 44 and 30 GHz. Pyramidal horns are either machined inside the
radiometer 20K module or connected via external electro-formed bended
waveguides. The requirement of high stability of the reference signal imposed a
careful design for the radiometric and thermal properties of the loads.
Materials used for the manufacturing have been characterized for thermal, RF
and mechanical properties. We describe in this paper the design and the
performance of the reference system.Comment: This is an author-created, un-copyedited version of an article
accepted for publication in JINST. IOP Publishing Ltd is not responsible for
any errors or omissions in this version of the manuscript or any version
derived from it. The definitive publisher authenticated version is available
online at [10.1088/1748-0221/4/12/T12006]. 14 pages, 34 figure
- …