Quantitative thermal imaging using single-pixel Si APD and MEMS mirror
Accurate quantitative temperature measurements are difficult to achieve using
focal-plane array sensors. This is due to reflections inside the instrument and the difficulty of
calibrating a matrix of pixels as identical radiation thermometers. Size-of-source effect (SSE),
which is the dependence of an infrared temperature measurement on the area surrounding the
target area, is a major contributor to this problem and cannot be reduced using glare stops.
Measurements are affected by power received from outside the field-of-view (FOV), leading
to increased measurement uncertainty. In this work, we present a micromechanical systems
(MEMS) mirror based scanning thermal imaging camera with reduced measurement
uncertainty compared to focal-plane array based systems. We demonstrate our flexible
imaging approach using a Si avalanche photodiode (APD), which utilises high internal gain to
enable the measurement of lower target temperatures with an effective wavelength of 1 µm
and compare results with a Si photodiode. We compare measurements from our APD thermal
imaging instrument against a commercial bolometer based focal-plane array camera. Our
scanning approach results in a reduction in SSE related temperature error by 66 °C for the
measurement of a spatially uniform 800 °C target when the target aperture diameter is
increased from 10 to 20 mm. We also find that our APD instrument is capable of measuring
target temperatures below 700 °C, over these near infrared wavelengths, with D* related
measurement uncertainty of ± 0.5 °C